• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. CONQUEST 40 PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. CONQUEST 40 PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number CQF7594
Medical Device Problem Codes Material Frayed (1262); Difficult to Remove (1528); Material Rupture (1546); Unraveled Material (1664)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/02/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date: 08/2023).
 
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon allegedly ruptured at 30 atm.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one conquest 40 pta dilatation catheter has been received for the evaluation.On the visual evaluation, the device was noted to be bloody.Frayed fibers and unraveling was noted at the proximal end of the balloon.No other specific anomalies were noted.On the functional evaluation, an in house presto inflation sample was used to inflate the sample and water was noted to be leaking at the proximal end of balloon.The balloon fibers were stripped and a longitudinal rupture was noted to the balloon at the distal tip.No evidence was noted for the difficult to remove from the evaluation.No other functional testing was performed.Therefore the investigation for the reported balloon rupture was confirmed as the water was leaking from the longitudinal ruptured region of the balloon.The investigation for the identified frayed and unraveled material was confirmed as the frayed and unraveling of the fibers were noted at the proximal end of the balloon.The investigation for the reported difficult to remove was inconclusive as the reported event cannot be replicated in the lab.A definitive root cause for the reported balloon rupture ,difficult to remove and the identified frayed, unraveled material could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: b5, d4 (expiry date: 08/2023),.
 
Event or Problem Description
It was reported that during an angioplasty procedure through the left arm, the pta balloon allegedly ruptured at 30 atm.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST 40 PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key12082303
Report Number2020394-2021-01257
Device Sequence Number16981414
Product Code DQY
UDI-Device Identifier00801741060601
UDI-Public(01)00801741060601
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K120660
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCQF7594
Device Catalogue NumberCQF7594
Device Lot NumberREEW1715
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/10/2021
Initial Date Received by Manufacturer 06/02/2021
Supplement Date Received by Manufacturer08/30/2021
Initial Report FDA Received Date06/29/2021
Supplement Report FDA Received Date09/01/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-