| Model Number |
72201919 |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/07/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a surgery the camera control unit when plugged in, the image was flickering.The procedure was successfully completed without delay using a back-up device.No patient injury or other complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device, used in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.The complaint of flickering could not be confirmed.Mdu passed functional testing and all functions perform as expected.Flickering did not occur during functional testing.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.
|
| |
|