|
Model Number 21-7302-24 |
Device Problem
Device Alarm System (1012)
|
Patient Problems
Headache (1880); Nausea (1970)
|
Event Date 05/24/2021 |
Event Type
malfunction
|
Event Description
|
Information was received indicating that while in use of a smiths medical cadd cassette reservoir, a no disposable, clamp tubing alarm was noted.It was also reported that the patient complained of headache, nausea, and stomachache.Subsequently, the cassette was changed, and alarm was cleared.No further adverse effects were reported.
|
|
Manufacturer Narrative
|
H10: device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.Samples received: the sample consist in one (1) cassette product from part number 21-7302-24 and lot number 4063420; the returned sample was received in used conditions decontaminated and inside in a plastic bag.Visual inspection: the sample was visually inspected at a distance of 12? to 16? under normal conditions of illumination to detect samples conditions that could cause functional issues.Results: the sample didn?t present any damage, scuffs, pinch marks, cracks, crazing, etc.Could cause the failure mode reported.Functional testing: sample was filled with water; the sample was connected to the cadd solis vip to look for unusual function.Results: the sample was fully priming and connected without difficult, the pump was set running and the alarm was not activated.The complaint was not confirmed.No actions taken were performed since the complaint was not confirmed.The cause of the reported problem could not be determined.
|
|
Search Alerts/Recalls
|
|
|