| Catalog Number |
03P84-25 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/22/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Apoc incident# (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
|
| |
|
Event or Problem Description
|
|
On 25-jun-2021, abbott point of care was contacted by a customer regarding i-stat crea cartridges that yielded a suspected discrepant creatinine result of 2.9 on an 84 year old female patient presented for a study scan and angiogram.There was no additional patient information available at the time of this report.Return product is available for investigation.Method:i-stat ; result : 2.9; date : (b)(6) 2021; time: 15:00.Lab , 0.9 , (b)(6) 2021 , 23:32.There are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident: # (b)(4).The investigation was completed on 26-jul-2021.A review of the device history record confirmed the lot passed finished goods release criteria.Retained and returned cartridge testing met the applicable acceptance criteria found in q04.01.003 rev.Ag, appendix 1- product complaint level 2 and level 3 investigation procedure.No deficiency has been identified for crea cartridge lot a21134.
|
| |
|
Event or Problem Description
|
|
Na.
|
| |
|