MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC OTHER EXTRA-DISCAL; SCREWDRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC OTHER EXTRA-DISCAL; SCREWDRIVER Back to Search Results
Model Number 815-517
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/03/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from healthcare professional via field representative regarding patient with symptoms of stenosis and radiculopathy involved in revision of past implants.It was reported that during procedure, both trial handle and the trial were stripped, they would not separate.Product came in contact with patient.There was no delay in overall procedure time.Product was utilized correctly according to the directions given in the ifu/labeling.There were no patient symptoms or complications reported as a result of this event.Patient is alive and no injury.
 
Event or Problem Description
Information was received from healthcare professional via field representative regarding patient with symptoms of stenosis and radic ulopathy involved in revision of past implants.It was reported that during procedure, both trial handle and the trial were stripped, they would not separate.Product came in contact with patient.There was no delay in overall procedure time.Product was utilized correctly according to the directions given in the ifu/labeling.There were no patient symptoms or complications reported as a result of this event.Patient is alive and no injury.On (b)(6) 2021, received additional information that in the reported event t10 to pelvis revision was performed.Drivers stripped during removals.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
H3: device evaluated result- visual inspection confirmed that the tip has been broken off, consistent with interface during usage.Inspection of the material hardness confirmed conformance to print specification.Optical examination of the fracture surface identified a fairly flat fracture surface and circular material displacement.This type of damage is consistent with torsional overload.Additional information- d9, g3, h3, h6.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OTHER EXTRA-DISCAL
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer Contact
david gustafson
1800 pyramid place
memphis, TN 38132
7635149628
MDR Report Key12085910
Report Number1030489-2021-00862
Device Sequence Number11851873
Product Code HXX
UDI-Device Identifier00673978045085
UDI-Public00673978045085
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number815-517
Device Catalogue Number815-517
Device Lot NumberKH18E450
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/03/2021
Supplement Date Received by Manufacturer07/01/2021
11/13/2021
Initial Report FDA Received Date06/29/2021
Supplement Report FDA Received Date07/21/2021
12/06/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/17/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexMale
-
-