STRYKER ENDOSCOPY-SAN JOSE SERFAS 50-S SWEEP + XL; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 0279351653
Medical Device Problem Code Continuous Firing (1123)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/05/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that the device was continuously activating.
 
Event or Problem Description
It was reported that the device was continuously activating.
 
Additional Manufacturer Narrative
Alleged failure: the ablation wand continued to stay active.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be an internal leak due to a broken/damaged bond of the polyimide and suction tubing causing fluid ingress to the pcb which can lead to a short circuit.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERFAS 50-S SWEEP + XL
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key12086362
Report Number0002936485-2021-00353
Device Sequence Number10984746
Product Code GEI
UDI-Device Identifier07613327405828
UDI-Public07613327405828
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K171391
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0279351653
Device Catalogue Number0279351653
Device Lot Number19262AE2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2021
Initial Date Received by Manufacturer 06/07/2021
Supplement Date Received by Manufacturer06/07/2021
Initial Report FDA Received Date06/29/2021
Supplement Report FDA Received Date08/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-