| Model Number |
0279351653 |
| Medical Device Problem Code |
Continuous Firing (1123)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/05/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that the device was continuously activating.
|
| |
|
Event or Problem Description
|
|
It was reported that the device was continuously activating.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: the ablation wand continued to stay active.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be an internal leak due to a broken/damaged bond of the polyimide and suction tubing causing fluid ingress to the pcb which can lead to a short circuit.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|