SMITH & NEPHEW, INC. VIS ADPT GUIDE KIT JII; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number V0100112
Medical Device Problem Code Fitting Problem (2183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/10/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during tka surgery the doctor pinned the visionaire femoral guide into place and made his distal femoral cut.When he put the size 4 planned size cutting guide into the pre drilled holes it seemed to have put them into extreme external rotation and would have taken off and estimated 15+ mm of bone off the posterior medial side.He then decided to change to conventional instrumentation and sized the femoral to a size 5 and the rotation was put correctly to 3 degrees of external rotation.They then used the visionaire tibial cutting guide and seemed to line up correctly and made the resection.When they checked the pre drilled tibial rotation, it again was extremely externally rotated.We then corrected that inter-op and sized it up one size from a 2 to a 3.Patient was not harmed as consequence of this problem.It is unknown if there was a surgical delay.
 
Additional Manufacturer Narrative
After further assessment of the information gathered by the manufacturer, it was identified that this event should be re-evaluated for mdr reporting.The original information stated that, during tka surgery, the doctor pinned the visionaire femoral guide into place and made his distal femoral cut.When he put the size 4 planned size cutting guide into the pre-drilled holes it seemed to have put them into extreme external rotation and would have taken off and estimated 15+ mm of bone off the posterior medial side.He then decided to change to conventional instrumentation and sized the femoral to a size 5 and the rotation was put correctly to 3 degrees of external rotation.They then used the visionaire tibial cutting guide and seemed to line up correctly and made the resection.When they checked the pre-drilled tibial rotation, it again was extremely externally rotated.We then corrected that inter-op and sized it up one size from a 2 to a 3.The procedure was completed with a delay of less than 30 minutes.Patient was not harmed as consequence of this malfunction.However, after further clarification and information received, it was determined that the present complaint does not meet the criteria to be reportable nor to be a complaint, because the two devices involved: the femoral and the tibial cutting blocks, are included in the vis cut guide jbcs kit rt.Therefore, the report already submitted (report number: 1020279-2021-05537) covers the malfunction of both devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIS ADPT GUIDE KIT JII
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12086387
Report Number1020279-2021-05535
Device Sequence Number11594248
Product Code JWH
UDI-Device Identifier00885556656419
UDI-Public00885556656419
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K170282
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberV0100112
Device Catalogue NumberV0100112
Device Lot Number00202490V2
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/14/2021
Supplement Date Received by Manufacturer06/14/2021
Initial Report FDA Received Date06/29/2021
Supplement Report FDA Received Date10/01/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-