Investigation: visual inspection: a slight deformation of the outer housing of the progav valve was observed through the visual inspection.The progav valve housing was subsequently measured and confirmed/indicated the presence of a deformation.The housing deformation measured at -0.40 mm, outside the tolerance of 0 ± 0.02 mm.Permeability test: a permeability test has shown that the valve is permeable.Adjustment test: the progav was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function did not operate as expected.The results indicate that the rotor no longer performs as required.Result: first, we performed a visual inspection of the progav.A deformation of the outer housing of the progav valve was observed through the visual inspection.This deformation was confirmed through a measurement of the plan parallelity.Next, we tested the permeability, adjustability, as well as the brake functionality.The brake function did not operate as expected, but all other specifications were met.Finally, we have dismantled the valve.Inside the valve we have found a build-up of substances (likely protein).Based on our investigation, we can determine that the progav had a brake malfunction at the time of our investigation.This is likely due to a deformation of the outer housing of the progav valve and due to the deposits observed inside the valve.Significant outside pressure, for example by too much force from the progav adjustment tool or by a fall or impact to the head of the patient, can compromise the integrity of the valve.As described in scientific literature, deposits caused by natural substances present in the liquor, such as protein, blood or tissue particles, are among the known and inevitable risks of hc-therapy by shunt implants.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|