| Model Number |
391239 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/08/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Combination product: yes.
|
| |
|
Event or Problem Description
|
|
An orsiro drug-eluting stent system was selected for treatment.During unpacking a stent deformation was noticed.Another device was used to proceed with the intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.The returned product was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation of the returned instrument revealed a deformation 12 mm distal to the proximal end of the stent, i.E.Struts are bent outwards and a mild kink.Microscopic analysis revealed stent imprints on the exposed balloon surface, indicating that the bent struts were initially crimped on the balloon.A reference protector could be sled over the deformed zone.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.As a part of final inspection every stent system undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent.Based on the conducted investigations, no manufacturing or material related root cause could be determined.
|
| |
|