Model Number 0998-00-0800-53
Device Problem Failure to Calibrate (2440)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/03/2021
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device (10/213): a getinge service territory manager (stm) was dispatched to investigate.The stm evaluated the iabp unit but was unable to duplicate the customer's reported issue.It was noted that there were no critical faults in the iabp log files.Subsequently, the stm completed scheduled preventative maintenance (pm) then completed all safety, functionality, and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.A supplemental report will be submitted if additional information is provided.The full event site name is (b)(6).
 
Event Description
It was reported that the cardiosave intra-aortic balloon pump (iabp) would not calibrate.It is unknown under which circumstances the event occurred or if a patient was involved; however there was no adverse event reported.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12088431
MDR Text Key259722845
Report Number2249723-2021-01362
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 09/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/03/2021
Initial Date FDA Received06/29/2021
Supplement Dates Manufacturer Received09/22/2022
Supplement Dates FDA Received09/27/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/27/2018
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-