| Model Number |
9-ITV09F45/80 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Arrhythmia (1721); Cardiac Tamponade (2226); Cardiac Perforation (2513); Pericardial Effusion (3271)
|
| Date of Event |
06/04/2021
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
|
| |
|
Event or Problem Description
|
|
It was reported that a 14mm amplatzer muscular ventricular septal defect (vsd occluder was selected for implant using a 9f amplatzer torqvue delivery system for a high risk, post myocardial infraction (mi) vsd closure on (b)(6) 2021.The (b)(6) year old patient had a vsd that was measured via computerized tomography (ct), transesophageal echocardiogram (tee), and left ventricular (lv) angiography.The defect measurement was about 10-11mm.The delivery system was placed in the internal jugular (ij) access site.Prior to deployment, a small pericardial effusion and a drop in the patient's pressures were observed on echocardiogram (echo).The device was attempted to be placed, however was deemed too small.The patient developed tamponade and the occluder was re-sheathed and removed from the patient.Pericardiocentesis was performed.It was believed that the pericardial effusion occurred while gaining access and crossing the defect.The patient's rhythm changed and became unstable.It was noted that the patient was shocked several times.An 18mm amplatzer post-infarct muscular vsd occluder was then attempted to be implanted using the same delivery system and placed appropriately within the defect; confirmed by tee.The patient continued to code for one hour and was declared deceased.The 18mm device was never released from the cable and not implanted.It was noted that the physician did not believe death was related to the vsd devices, instead suspected tamponade was caused by cardiac perforation prior to device deployment.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of pericardial effusion, drop in "pressures", tamponade, arrhythmia, and patient death was reported.A more comprehensive assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Search Alerts/Recalls
|