• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) OCTRODE LEAD KIT, 60CM LENGTH; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) OCTRODE LEAD KIT, 60CM LENGTH; SCS LEAD Back to Search Results
Model Number 3186
Device Problems High impedance (1291); Material Twisted/Bent (2981)
Patient Problem Inadequate Pain Relief (2388)
Event Date 05/14/2021
Event Type  Injury  
Manufacturer Narrative
Date of event is estimated.
 
Event Description
Related manufacturer reference number: 3006705815-2021-03181.It was reported that the patient's lead was displaying high impedances.Reprogramming was unable to resolve the issue.Surgical intervention is anticipated.It is unknown which lead is displaying high impedances so both are being reported on.
 
Event Description
It was reported that the physician explanted and replaced the patient's leads.Reportedly, one lead was displaying high impedances and the other was bent.Surgical intervention resolved the issue.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE LEAD KIT, 60CM LENGTH
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
MDR Report Key12088667
MDR Text Key259127493
Report Number3006705815-2021-03182
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067017246
UDI-Public05415067017246
Combination Product (y/n)N
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 07/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/29/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/25/2021
Device Model Number3186
Device Catalogue Number3186
Device Lot NumberA000090596
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
3186
Patient Outcome(s) Other;
Patient Weight84
-
-