The reported event was confirmed however the cause was unknown.Visual inspection noted one unopened magic 3 intermittent catheter was received.The water packet appeared to be flat and empty prior to opening the catheter package.There was a evidence that the packet was popped or water residue inside the package.Although an exact root cause could not be determined however a potential root cause for this failure mode could be due to misaligned film during the folding process.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "intended use: the catheter is intended for urinary bladder drainage in patients requiring catheterization for management of incontinence, voiding dysfunction, and surgical procedures.Please contact your physician to determine which product options are best for you, paying close attention to product warnings or precautions and adverse reactions.The catheter becomes slippery when wetted with water, eliminating the need for a separate lubricant.For your added convenience, this catheter is packaged with its own sterile water.Simply release the water from its foil packet and then tip the unopened catheter package end to end.The catheter acts like a magnet to attract the water and activate its slippery coating.Directions for use: 1.Release the water.Prior to opening the sealed catheter pouch: a.Apply pressure to the foil packet to release the water.B.Ensure all water is released from the foil packet.2.Wet the catheter.A.Hold package with printed side up.B.Tip package end-to-end three to six times to wet catheter.This movement is required so that the water transfers back and forth over the catheter to fully wet the hydrophilic coating.3.Use the catheter a.Peel back package to expose funnel end of catheter.B.If desired, use the self-adhesive tape on the package to temporarily attach the pouch to any dry vertical surface.C.Remove catheter and use according to physicians instructions.Warning: this is a single use device.Do not reuse.Reuse of a single use device increases the risk of catheter acquired urinary tract infections.Urethral catheter for urological use only.Discard after use, do not resterilize.Made of silicone elastomer." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|