• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG SOLUTIONS S.A. MAHURKAR; CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG SOLUTIONS S.A. MAHURKAR; CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Model Number 8813793013
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Unintended Radiation Exposure (4565)
Date of Event 05/25/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during the removal of the guide wire after passage of the catheter, the guide wire broke into two (guide wire was unwound).The catheter was not repaired and there was no leak.Alcoholic chlorhexhydrin degerant was used as the cleaning agent on the device and was also used at the insertion site prior to product placement.There was no excessive force used to pull/remove the product and there was nothing unusual observed on the device before use.Before the breakage, there were no abnormal conditions noted.Flushing was done which had a normal result.There were no other products utilized with the device and no other defects/damages found on the product.The guide wire used was the one included in the kit.There were no tools used when they tried to remove the guide wire.An x-ray machine was used to check for retained parts.All parts were recovered.The product had to be replaced with a new one to resolve the issue and the procedure was completed.There was no blood loss and no intervention/treatment was required as a result of the event.It did not lead to or prolonged the patient's hospitalization.
 
Additional Manufacturer Narrative
Correction: g1 (email, phone number).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAHURKAR
Common Device Name
CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
CS  20101
MDR Report Key12089926
Report Number3009211636-2021-00152
Device Sequence Number14450176
Product Code FOZ
UDI-Device Identifier10884521004832
UDI-Public10884521004832
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K943349
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/13/2022
Device Model Number8813793013
Device Catalogue Number8813793013
Device Lot Number1801000112
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/24/2021
Supplement Date Received by Manufacturer06/24/2021
Initial Report FDA Received Date06/30/2021
Supplement Report FDA Received Date06/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age38 YR
-
-