| Model Number |
INX500T15 |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Codes |
Dyspnea (1816); Cough (4457); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/19/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A continuous positive airway pressure (cpap) device was returned to a third-party service center for service.There was no report of patient harm or injury.During the evaluation of the device at a third-party service center, the sound abatement foam was found to be degraded.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer previously reported an allegation related to the device's sound abate foam.This action was reported to fda per 21 cf1 part 806.The device will be corrected per res 88058.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer previously received information alleging an issue related to a cpap device's sound abatement foam.The customer had alleged to have black particles blowing out of the device.The manufacturer reported there was no report of patient harm or injury.In the initial report, the patient's allegation of having a dry cough and shortness of breath was inadvertently omitted.There was no report of serious patient harm or injury.There was no report of medical intervention.The manufacturer has corrected the health effect clinical code and health effect impact code in this report.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer previously received information alleging an issue related to a cpap device's sound abatement foam.The customer had alleged to have black particles blowing out of the device.There was no report of patient harm or injury.The device was returned to a third-party service center for investigation.During the evaluation of the device, the third-party service center visually inspected the device and did not note any finding of sound abatement foam degradation.The device did not pass final testing.The device was scrapped.The manufacturer has updated h6 in this report.
|
| |
|
Search Alerts/Recalls
|