MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION AED PRO; DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION AED PRO; DEFIBRILLATOR Back to Search Results
Model Number AED PRO
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event or Problem Description
Complainant alleged that during functional testing, the device would not power up.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Additional Manufacturer Narrative
The device was returned to zoll medical corporation; the customer's report was observed and attributed to a system interconnect flex cable that was not properly seated on the analog board.The cable was reseated to correct the reported problem.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AED PRO
Common Device Name
DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key12093112
Report Number1220908-2021-02057
Device Sequence Number13687199
Product Code MKJ
UDI-Device Identifier00847946016135
UDI-Public00847946016135
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K110526
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberAED PRO
Device Catalogue NumberAED PRO
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/09/2021
Supplement Date Received by Manufacturer06/09/2021
Initial Report FDA Received Date06/30/2021
Supplement Report FDA Received Date07/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-