• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US BIOKNOTLESS RAPIDE W/PAN; SOFT-TISSUE ANCHOR, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US BIOKNOTLESS RAPIDE W/PAN; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Model Number 212723
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Udi: (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported by the sales rep in (b)(6) that during an unknown surgery on an unknown date, it was observed that the bioknotless rapide w/pan device broke above the tip.There were no adverse patient consequences reported.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: b5: subsequent follow-up with the customer, additional information was received regarding the event.Therefore, the event description has been updated accordingly.
 
Event or Problem Description
It was reported by the sales rep in united kingdom that during an instability repair procedure on an unknown date, it was observed that the bioknotless rapide w/pan device broke above the tip.There were no fragments left in the patient's body.Another like device was used to complete the procedure.There were no adverse patient consequences reported.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch ill be filed as appropriate.H10 additional narrative: investigation summary: according to the information provided, it was reported by the sales rep via email that during an unknown procedure the bioknotless br anchor w/panacryl broke above the anchor tip.Since the complaint device was not returned, we cannot determine a root cause for the reported failure.If the device is received in the future, we will reopen the complaint and perform the investigation as appropriate.A manufacturing record evaluation was performed for the finished device lot number: 6l87435, and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOKNOTLESS RAPIDE W/PAN
Common Device Name
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key12093267
Report Number1221934-2021-02043
Device Sequence Number3979732
Product Code MAI
UDI-Device Identifier10886705002092
UDI-Public10886705002092
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K150209
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2023
Device Model Number212723
Device Catalogue Number212723
Device Lot Number6L87435
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/16/2021
Supplement Date Received by Manufacturer07/09/2021
08/19/2021
Initial Report FDA Received Date06/30/2021
Supplement Report FDA Received Date07/12/2021
08/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-