• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. TURON SHOULDER; TURON PEGGED GLENOID IMPLANT, SIZE 50

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. TURON SHOULDER; TURON PEGGED GLENOID IMPLANT, SIZE 50 Back to Search Results
Model Number 520-01-250
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/15/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event or Problem Description
Revision surgery - due to unknown reason.
 
Event or Problem Description
Revision surgery - patient needed to be revised to a reverse.
 
Additional Manufacturer Narrative
Corrected data: see b.2., b.5.& d.4.Manufacturer narrative: the reason for this revision surgery was reported as revise to a reverse.The previous surgery and the surgery detailed in this event occurred 7.9 years apart.The healthcare professional indicated there was a significant adverse event to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery to revise a reverse.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURON SHOULDER
Common Device Name
TURON PEGGED GLENOID IMPLANT, SIZE 50
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
MDR Report Key12093302
Report Number1644408-2021-00652
Device Sequence Number17208335
Product Code KWS
UDI-Device Identifier00888912024976
UDI-Public(01)00888912024976
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K123982
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/28/2017
Device Model Number520-01-250
Device Catalogue Number520-01-250
Device Lot Number193G1047
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/15/2021
Supplement Date Received by Manufacturer07/26/2021
Initial Report FDA Received Date06/30/2021
Supplement Report FDA Received Date08/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
520-00-000 LOT 878C1231; 520-46-024 LOT 54005601
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age55 YR
-
-