Investigation: visual inspection: no significant deformations or damage of the progav 2.0 were detected during the visual inspection.Permeability test: a permeability test has shown that the progav 2.0 is permeable.Adjustment test: the progav 2.0 was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test of the progav 2.0 has shown that the brake function is fully operational and the braking force is within the given tolerances.Result: first, we performed a visual inspection of the progav 2.0.No significant deformations or damage of the valve were detected during the visual inspection.Next, we tested the permeability, adjustability, as well as the brake functionality and brake force.The valve operates as expected and met all specifications.Finally, we have dismantled the valve.Inside the valve we have found a build-up of substances (likely protein).Based on our investigation, we are unable to substantiate the claim of non-adjustability.At the time of our investigation, the progav 2.0 was able to be adjusted to all specified settings.However, it is possible that the deposits observed inside the valve could have caused the malfunction in the past.As described in scientific literature, the problem encountered is one of the known, inevitable risks of hc-therapy by shunt implants.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke (b)(4).No further regulatory actions are required from our point of view.
|