| Model Number |
24704 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/30/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: returned product consisted of a coyote balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed no damages.Microscopic examination revealed a pinhole in the balloon 8mm from the tip.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 10june2021.It was reported that balloon inflation failure occurred.The target lesion was located in the lower extremity vessel.A 2.0mm x 60mm x 150cm coyote balloon catheter was advanced for dilation but the balloon would not inflate.The device was removed and the procedure was completed with another of the same device.There were no patient complications reported.However, returned device analysis revealed balloon pinhole.
|
| |
|