• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: K2M, INC. TAPERED DRIVER; SIZE 10; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

K2M, INC. TAPERED DRIVER; SIZE 10; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number 201-90118
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/07/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that 2- pyrenees tapered driver tips were stripped intra-operatively.The procedure was completed successfully without surgical delay.No adverse consequences or medical intervention were reported.This report captures the second of two drivers.
 
Additional Manufacturer Narrative
Visual inspection: it was observed that the hexalobe tip was deformed.The edges of the hexalobe tip were deformed in a rotational pattern and appeared dull.Device history records were reviewed for this lot, and no relevant manufacturing issues were identified.Complaint history records were reviewed for this lot, and five similar complaints were identified.Manufacturing records reveal that the device was about six years old at the time of event, as it was manufactured in 2015.It is likely that repeated use during the device's service life compromised the material integrity, causing tip deformation.
 
Event or Problem Description
It was reported that 2- pyrenees tapered driver tips were stripped intra-operatively.The procedure was completed successfully without surgical delay.No adverse consequences or medical intervention were reported.This report captures the second of two drivers.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TAPERED DRIVER; SIZE 10
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
K2M, INC.
600 hope parkway se
leesburg VA 20175
MDR Report Key12094526
Report Number3004774118-2021-00197
Device Sequence Number4000577
Product Code LXH
UDI-Device Identifier10888857030183
UDI-Public10888857030183
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number201-90118
Device Lot NumberATCC
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/08/2021
Initial Date Received by Manufacturer 06/07/2021
Supplement Date Received by Manufacturer09/30/2021
Initial Report FDA Received Date06/30/2021
Supplement Report FDA Received Date10/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-