• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION LINEAR ST; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION LINEAR ST; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number SC-2218-50
Device Problems High impedance (1291); Migration (4003); Intermittent Energy Output (4025)
Patient Problems Headache (1880); Nausea (1970); Inadequate Pain Relief (2388)
Event Date 02/20/2020
Event Type  Injury  
Manufacturer Narrative
Date of event: approximated based on the date the manufacturer became aware of the event.Additional suspect medical device component involved in the event: product family: scs-linear leads upn: (b)(4).Model: sc-2218-50 serial: (b)(4).Batch: 7073649.
 
Event Description
It was reported that the patient experienced loss of stimulation due to lead migration.The stimulation goes off and on when walking or making posture changes.Positional headaches and nausea were also noted.Database analysis of the charge profile revealed no anomalies.The impedance measurements revealed high values on port d(1 contact).The patient only gets left ribs and abdomen stimulation.Reprogramming attempt was done with no resolution.All components were explanted and will not be returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINEAR ST
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key12095461
MDR Text Key260744970
Report Number3006630150-2021-03367
Device Sequence Number1
Product Code LGW
UDI-Device Identifier08714729767725
UDI-Public08714729767725
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 06/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/30/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/12/2021
Device Model NumberSC-2218-50
Device Catalogue NumberSC-2218-50
Device Lot Number7073565
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/10/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/17/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;