TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX25RWC W/ RES; BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3CX*FX25RWC
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/11/2021
Event Type  malfunction  
Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during prime, the customer discovered a damaged / cracked oxygenator.No patient involvement.The product was changed out.Procedure was completed successfully.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added exp date) g3 (date received by manufacturer) g6 (indication that this is a follow-up report) h2 (follow-up due to additional information and device evaluation) h3 (device evaluated by manufacturer) h4 (device manufacture date) h6 (identification of evaluation codes 10, 11, 3331, 3259, 19) type of investigation #1: 10 - testing of actual/suspected device type of investigation #2: 11 - testing of device from same lot/batch retained by manufacturer type of investigation #3: 3331 - analysis of production records investigation findings: 3259 - improper physical structure investigation conclusions: 19 - cause traced to user the affected sample was inspected upon receipt to confirm a mark on the oxygenator blood inlet.The sample was setup in a water circuit and circulated at approximately 1000mmhg for approximately 10 minutes with no leakage observed.A representative retention sample was inspected for damage to the blood inlet with no damage noted.The mark resembles a cut from a knife to cut tubing off of the port, however, there are no knives within the production area and tubing is not added or removed within production.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on june 30, 2021.  upon further investigation of the reported event, the following information is new and/or changed: d9 (device availability - added date returned to manufacturer).G3 (date received by manufacturer).G6 (indication that this is a follow-up report).H2 (follow-up due to additional information).H3 (device evaluation anticipated by manufacturer - a second follow-up will be submitted upon completion of the investigation and/or submission of new information, thus tcvs references conclusions code 11.) all available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERILE FX25RWC W/ RES
Type of Device
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
MDR Report Key12096088
MDR Text Key261540621
Report Number1124841-2021-00161
Device Sequence Number1
Product Code DTZ
UDI-Device Identifier00699753450837
UDI-Public(01)00699753450837
Combination Product (y/n)N
PMA/PMN Number
K151791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 08/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2023
Device Model Number3CX*FX25RWC
Device Catalogue NumberN/A
Device Lot NumberYK10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/30/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/15/2021
Initial Date FDA Received06/30/2021
Supplement Dates Manufacturer Received06/30/2021
08/18/2021
Supplement Dates FDA Received07/16/2021
08/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-