Model Number 3CX*FX25RWC |
Device Problem
Crack (1135)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/11/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
|
|
Event Description
|
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during prime, the customer discovered a damaged / cracked oxygenator.No patient involvement.The product was changed out.Procedure was completed successfully.
|
|
Manufacturer Narrative
|
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added exp date) g3 (date received by manufacturer) g6 (indication that this is a follow-up report) h2 (follow-up due to additional information and device evaluation) h3 (device evaluated by manufacturer) h4 (device manufacture date) h6 (identification of evaluation codes 10, 11, 3331, 3259, 19) type of investigation #1: 10 - testing of actual/suspected device type of investigation #2: 11 - testing of device from same lot/batch retained by manufacturer type of investigation #3: 3331 - analysis of production records investigation findings: 3259 - improper physical structure investigation conclusions: 19 - cause traced to user the affected sample was inspected upon receipt to confirm a mark on the oxygenator blood inlet.The sample was setup in a water circuit and circulated at approximately 1000mmhg for approximately 10 minutes with no leakage observed.A representative retention sample was inspected for damage to the blood inlet with no damage noted.The mark resembles a cut from a knife to cut tubing off of the port, however, there are no knives within the production area and tubing is not added or removed within production.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|
|
Manufacturer Narrative
|
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on june 30, 2021. upon further investigation of the reported event, the following information is new and/or changed: d9 (device availability - added date returned to manufacturer).G3 (date received by manufacturer).G6 (indication that this is a follow-up report).H2 (follow-up due to additional information).H3 (device evaluation anticipated by manufacturer - a second follow-up will be submitted upon completion of the investigation and/or submission of new information, thus tcvs references conclusions code 11.) all available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|
|