| Model Number |
FRED3009-PMA |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
Ischemia Stroke (4418)
|
| Date of Event |
05/25/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformances associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was implanted in the patient and is not available for return to the manufacturer for analysis; therefore, the alleged product issue cannot be confirmed.The instructions for use (ifu) identifies stroke or tia (transient ischemic attack), migration or misplacement as potential complications associated with use of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that after successfully implanting the fred 21 stent in the m1 segment of middle cerebral artery (mca) aneurysm, the patient came back 6 days post implantation with a stroke.Angiogram run showed the stent migrated down the m1 vessel blocking the bifurcation.Intervention was taken and the patient was treated with direct catheter integrilin and drip.
|
| |
|