SAFESTEP HUBER NEEDLE SET 20G X 1 IN WITH Y-SITE; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number N/A
Medical Device Problem Code Component Missing (2306)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/07/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of asdps0165 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported that when opening the package, it was noticed that the cover of needle was not contained in it.There was no reported patient involvement.No other information provided.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, complaint and lot history review, applicable previous investigation(s), sample (if available), labeling, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a missing needle cover is inconclusive due to the state of the returned sample.One 22 g x 1 in.Safestep infusion set with a valved y-site was returned for evaluation.An initial visual observation showed no obvious evidence of use on the returned sample.A cap was returned on the proximal luer hub, but no needle guard was returned on or with the returned sample.While no needle guard was returned on or with the returned sample, the sample was returned outside of any packaging which made it difficult to determine when and where the component may have gone missing.
 
Event or Problem Description
It was reported that when opening the package, it was noticed that the cover of needle was not contained in it.There was no reported patient involvement.No other information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFESTEP HUBER NEEDLE SET 20G X 1 IN WITH Y-SITE
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
MDR Report Key12100117
Report Number3006260740-2021-02528
Device Sequence Number3652223
Product Code FPA
UDI-Device Identifier00801741066191
UDI-Public(01)00801741066191
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K153440
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberLH-0030YN
Device Lot NumberASDPS0165
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2021
Initial Date Received by Manufacturer 06/08/2021
Supplement Date Received by Manufacturer07/19/2021
Initial Report FDA Received Date07/01/2021
Supplement Report FDA Received Date08/13/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/01/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-