| Model Number |
4485 |
| Medical Device Problem Codes |
Incorrect Measurement (1383); Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/08/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer reported that the cable provided "confused values" or an error message.No consequences or impact to patient were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: other, other text: the returned device was evaluated.No physical damage was observed during visual inspection.During continuity testing with a tester, there was an open in the cable on the green conductor, inside the bend relief area of the m15 connector.Testing with a monitor showed a "sensor off" error message displayed and an audible and visual alarm.The cable is malfunctioning and is unable to perform spo2 and pulse rate simulation tests.
|
| |
|
Event or Problem Description
|
|
The customer reported that the cable provided "confused values" or an error message.No consequences or impact to patient were reported.
|
| |
|