ZIMMER BIOMET, INC. ACT ARTIC E1 HIP BRG 28X44MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 06/08/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The device will not be returned for analysis, due to hospital protocol; however, an investigation of the reported event is in progress.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 001825034-2021-01950.
 
Event or Problem Description
It was reported that the patient underwent a revision procedure approximately 2 months post implantation due to dislocation.During the surgery, while converting a dm liner to a g7 freedom liner to address dislocation, the surgeon tried all recommended extraction techniques to remove the metal liner, but could not be removed.Liner was eventually removed with cut it out with a metal cutting burr.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device history records identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined as to why the metal liner and bearing dislocated.The root cause of the metal liner and shell not coming apart is the product design as the desired outcome of the two components is longevity and to be well connected.No failure of the devices.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACT ARTIC E1 HIP BRG 28X44MM
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key12102220
Report Number0001825034-2021-01951
Device Sequence Number3577591
Product Code LPH
UDI-Device Identifier00880304485051
UDI-Public(01)00880304485051(17)240906(10)199510
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K101336
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/06/2024
Device Model NumberN/A
Device Catalogue NumberEP-200150
Device Lot Number199510
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/10/2021
Supplement Date Received by Manufacturer10/26/2021
Initial Report FDA Received Date07/01/2021
Supplement Report FDA Received Date10/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CAT#: 110010246 G7 SHELL LOT#: 64809421; CAT#: 110024464 G7 LINER LOT#: 429260; UNKNOWN SHELL; CAT#: 110024464 G7 LINER LOT#: 429260; UNKNOWN SHELL
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-