|
Main investigation conclusion:
the reported event of a backup battery fault alarm was confirmed via the submitted log file.The log file contained approximately 5 days of data ((b)(6) 2021 ¿ (b)(6) 2021 per the timestamp).A backup battery fault alarm activated at 2:14:20 on (b)(6) 2021 due to a backup battery load test failure (error code f36).The backup battery failed the load test due to the loaded voltage being under the acceptable threshold compared to the unloaded voltage.The alarm did not affect the controller's ability to operate the pump at the set speed.The driveline was disconnected at 2:16:18 on (b)(6) 2021 and both power cables were disconnected at 2:16:58 to exchange the system controller.The controller was then put into sleep mode at 2:17:29 on (b)(6) 2021.The backup battery fault alarm remained active until the controller was put into sleep mode.The controller was then reconnected to the mobile power unit (mpu) at 2:17:57 and operated in charge mode with no alarms active.The driveline was then reconnected at 2:21:16 on (b)(6) 2021.The pump then started at 2:24:15 on (b)(6) 2021.The pump ramped up to the fixed speed by 2:24:16 and then remained above the low speed limit for the remainder of the log file.Additional information provided stated that the system controller, serial number (b)(6), would be returned for evaluation; however, the controller has not been received to date.Multiple requests for additional information were made to determine if the controller will still be returned and to obtain the tracking number for the shipment; however, no response was received.This file will be reopened at a later date if the controller is returned or a tracking number is provided.The root cause of the reported event could not be conclusively determined through this analysis.The heartmate 3 patient handbook and the heartmate 3 instructions for use (ifu) cover all alarms (visual and audible), including the backup battery fault alarm, and the actions to take when the alarms occur.When backup battery fault alarms occur, the patient is instructed to call their hospital contact immediately for diagnosis and instructions.Heartmate 3 instructions for use, rev.C, section 7 ¿ ¿alarms and troubleshooting¿ and heartmate 3 patient handbook, rev.C, section 5 ¿ ¿alarms and troubleshooting¿ covers all alarms (visual and audible), including backup battery fault and power cable disconnect alarms, and the actions to take if the alarm cannot be resolved.When a backup battery fault alarm occurs, the patient is instructed is instructed to call their hospital contact as soon as possible for diagnosis and instructions.The ifu and patient handbook do not instruct the patient to exchange the system controller in response to a backup battery fault alarm.Heartmate 3 patient handbook, rev.C, section 2 ¿how your heart pump works¿ explains how to replace the running system controller with a backup controller.This section instructs the user not to attempt to change the system controller without having a trained, competent caregiver by their side to assist.The user is instructed to follow all alarm instructions, including calling the hospital.Heartmate 3 instructions for use (ifu), rev.C, section 2 ¿system operations¿ also explains how to replace the running system controller with a backup controller, and instructs the user to ensure that patients understand the need for having a caregiver present during a controller exchange and that all labelling instructions are followed, including calling the hospital contact.Heartmate 3 patient handbook, rev.C, section 6 "caring for the equipment" and heartmate 3 instructions for use (ifu), rev.C, section 8 ¿equipment storage and care¿ describe how to care for and clean all equipment.Heartmate 3 patient handbook, rev.C, section 10 and heartmate 3 instructions for use (ifu), rev.C, section f, both entitled ¿safety checklists¿, provide checklists to assist the patient in performing routine maintenance of heartmate 3 lvad.Heartmate 3 patient handbook, rev.C, cautions users to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.Device history records were reviewed and showed no deviations from manufacturing or qa specifications.The system controller, serial number (b)(6), was shipped to the customer on (b)(6) 2020.No further information was provided.The manufacturer is closing the file on this event.
|