• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS CORPORATION UNKNOWN OPTEASE VENA CAVA FILTER; FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CORPORATION UNKNOWN OPTEASE VENA CAVA FILTER; FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number 466FXXXX
Device Problem Unintended Movement (3026)
Patient Problems Pain (1994); Unspecified Tissue Injury (4559)
Event Date 06/18/2021
Event Type  Injury  
Manufacturer Narrative
Reporter occupation: other, senior counsel, litigation.Please note that the exact event date is unknown and the event date is the complaint awareness date.It was reported that a patient underwent placement of an optease vena cava filter.The information provided indicated that the filter subsequently malfunctioned and caused tilt.The indication for the filter implant, procedural details and medical history of the patient have not been provided and there is currently no additional information available for review.The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed.The optease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Post implant imaging has not been provided.Without the procedural films or post-placement imaging and the limited information provided, the report of tilt could not be confirmed or further clarified.Additionally, the timing and mechanism of the filter tilt is unknown.Ivc filter tilt has been associated with operator technique, vessel characteristics, specifically asymmetry and tortuosity.Given the limited information currently available for review, there is nothing to suggest that the reported events are related to the design and manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.Please note that this is the initial report for this product.
 
Event Description
As reported by the legal brief, the patient underwent placement of an optease vena cava filter.The report states that the filter subsequently malfunctioned and caused injury and damage to the patient including, but not limited to tilt.As a direct and proximate result of these malfunctions, the patient suffered life-threatening injuries and damages, and required extensive medical care and treatment.As a further proximate result, the patient has suffered and will continue to suffer significant medical expenses, pain and suffering, and other damages.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN OPTEASE VENA CAVA FILTER
Type of Device
FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer (Section G)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer Contact
karla castro
14201 nw 60th ave
miami lakes, FL 33014
7863138372
MDR Report Key12104257
MDR Text Key259592344
Report Number1016427-2021-05034
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Other
Type of Report Initial
Report Date 07/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/01/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number466FXXXX
Was Device Available for Evaluation? No
Date Manufacturer Received06/18/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
-
-