| Model Number |
761165 |
| Medical Device Problem Codes |
Short Fill (1575); Volume Accuracy Problem (1675)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/04/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported when using the paxgene® blood dna tube, the device experienced underfill or low draw of a tube with blood.This event occurred 100 times.The following information was provided by the initial reporter.The customer stated: tubes don't fill correctly.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 2021-07-20.Investigation summary: bd received 13 samples for investigation.The samples were evaluated by functional draw testing and the indicated failure mode for underfill with the incident lot was not observed.Additionally, 7 retention samples from bd inventory were evaluated by functional draw testing and no issues were observed relating to underfill as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported when using the paxgene® blood dna tube, the device experienced underfill or low draw of a tube with blood.This event occurred 100 times.The following information was provided by the initial reporter.The customer stated: tubes don't fill correctly.
|
| |
|