Model Number EICA7070-01
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Adhesion(s) (1695)
Date of Event 05/26/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference case-(b)(4).Literature: low-temperature radiofrequency coblation reduces treatment interval and post-operative pain of laryngotracheal recurrent respiratory papillomatosis doi: 10.1042/bsr20192005.
 
Event or Problem Description
It was reported that on literature review ¿low-temperature radiofrequency coblation reduces treatment interval and post-operative pain of laryngotracheal recurrent respiratory papillomatosis.¿, three patients developed vocal cord adhesions after procedures in which a procise lw was used.No further information is available.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).H10 h3, h6: the reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.Insufficient product identification information was provided and thus a manufacturing record review could not be conducted.A complaint history review concluded this was an isolated event.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.A clinical review states the data provided was presented in the literature article named " low-temperature radiofrequency coblation reduces treatment interval and post-operative pain of laryngotracheal recurrent respiratory papillomatosis." without clinically relevant patient-specific supporting documentation, a thorough medical investigation could not be performed.The root cause and/or patient outcome beyond that which was documented in the article could not be confirmed nor concluded; therefore, no further medical assessment is warranted at this time.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCISE LW COBLATOR II
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key12104869
Report Number3006524618-2021-00732
Device Sequence Number3577630
Product Code GEI
UDI-Device Identifier00817470004243
UDI-Public00817470004243
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K022006
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberEICA7070-01
Device Catalogue NumberEICA7070-01
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/12/2021
Supplement Date Received by Manufacturer10/14/2021
Initial Report FDA Received Date07/01/2021
Supplement Report FDA Received Date10/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-