Model Number CD3265-40Q
Medical Device Problem Codes Premature Discharge of Battery (1057); Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/14/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.The device is included in the battery performance alert advisory issued by abbott on 28 august 2017.
 
Event or Problem Description
Following the battery performance alert (bpa) advisory, a bpa was received by the clinician and the device was explanted and replaced.The patient was stable.
 
Additional Manufacturer Narrative
Interrogation of the device revealed it was above eri when received.Based on this information, the device was found to communicate appropriately with a merlin programmer and has not reached the elective replacement indicator (eri).The device is included in the premature battery depletion with implantable cardioverter defibrillator advisory notice issued by st.Jude medical on 11 october 2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA (DDQP+)
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
MDR Report Key12104996
Report Number2017865-2021-24911
Device Sequence Number3800313
Product Code NIK
UDI-Device Identifier05414734505673
UDI-Public05414734505673
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P030054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 09/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2015
Device Model NumberCD3265-40Q
Device Catalogue NumberCD3265-40Q
Device Lot Number4149992
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/05/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/15/2021
Supplement Date Received by Manufacturer09/10/2021
Initial Report FDA Received Date07/01/2021
Supplement Report FDA Received Date09/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0004-2018
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age55 YR
-
-