• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION Back to Search Results
Model Number MS1-4590R
Medical Device Problem Codes Degraded (1153); Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 09/01/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to lirc for investigation.The device was not returned to the manufacturer for investigation.Root cause is unknown at this time.
 
Event or Problem Description
It was reported that a revision procedure was performed due to the rod having a pin fracture.During a review of investigation findings from the (b)(6) it was identified that the rod had a pin fracture with debris in the housing tube, and the presence of discoloration/corrosion was noted.
 
Additional Manufacturer Narrative
Additional information: no product has been returned for evaluation, however, investigation of the failure revealed that corrosion, surface finish and bending stresses all contribute to pin breaking.A new pin was implemented that had 65% improved strength and significantly improved corrosion resistance.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Common Device Name
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, suite 100
aliso viejo CA 92656
MDR Report Key12105068
Report Number3006179046-2021-00347
Device Sequence Number14606250
Product Code PGN
UDI-Device Identifier00812258026301
UDI-Public812258026301
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K171791
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2016
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 07/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMS1-4590R
Device Lot NumberA130508-01
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/16/2021
Supplement Date Received by Manufacturer07/27/2021
Initial Report FDA Received Date07/01/2021
Supplement Report FDA Received Date07/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age6 YR
-
-