| Model Number |
MS1-4590R |
| Medical Device Problem Codes |
Degraded (1153); Fracture (1260)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/01/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to lirc for investigation.The device was not returned to the manufacturer for investigation.Root cause is unknown at this time.
|
| |
|
Event or Problem Description
|
|
It was reported that a revision procedure was performed due to the rod having a pin fracture.During a review of investigation findings from the (b)(6) it was identified that the rod had a pin fracture with debris in the housing tube, and the presence of discoloration/corrosion was noted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: no product has been returned for evaluation, however, investigation of the failure revealed that corrosion, surface finish and bending stresses all contribute to pin breaking.A new pin was implemented that had 65% improved strength and significantly improved corrosion resistance.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|