Model Number EIC7070-01
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Scar Tissue (2060)
Date of Event 05/26/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference case (b)(4).Literature: cold-plasma posterior arytenoidochordectomy for chronic paralytic laryngostenosis.Doi: 10.31838/srp.2020.3.98.
 
Event or Problem Description
It was reported that on literature review ¿cold-plasma posterior arytenoidochordectomy for chronic paralytic laryngostenosis¿, after an ent procedure with a "procise lw coblator ii wand", a patient developed a scar membrane.The complication was treated with a second surgery to remove the scar.No further information is available.
 
Additional Manufacturer Narrative
H10: h3, h6: the reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.Insufficient product identification information was provided and thus a manufacturing record review could not be conducted.A complaint history review concluded this was an isolated event.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.A clinical review states the data provided was presented in the literature article named " cold-plasma posterior arytenoidochordectomy for chronic paralytic laryngostenosis." without clinically relevant patient-specific supporting documentation, a thorough medical investigation could not be performed.The images provided in the article have been interpreted within the text; therefore, no further analysis of the images is required.The root cause and/or patient outcome beyond that which was documented in the article could not be confirmed nor concluded; therefore, no further medical assessment is warranted at this time.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCISE LW COBLATOR II
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key12105076
Report Number3006524618-2021-00738
Device Sequence Number11775249
Product Code GEI
UDI-Device Identifier00817470000443
UDI-Public00817470000443
Combination Product (Y/N)N
Initial Reporter CountryRS
PMA/510(K) Number
K202006
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberEIC7070-01
Device Catalogue NumberEIC7070-01
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/12/2021
Supplement Date Received by Manufacturer10/14/2021
Initial Report FDA Received Date07/01/2021
Supplement Report FDA Received Date10/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age50 YR
-
-