MEDTRONIC, INC. CARELINK PROGRAMMER; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 2090W
Device Problems Computer Software Problem (1112); No Display/Image (1183); Device Difficult to Program or Calibrate (1496); Defective Device (2588); Computer Operating System Problem (2898)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/04/2021
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that stylus of the programmer was unable to be used.The stylus was replaced, however the issue persisted.The programmer displayed a black screen with a "calibrate stylus" message.It was further reported that the programmer displayed a "stylus is out of range" message, and that the programmer was in bad condition the programmer was unable to boot up.It was also reported that the issue did not resolve after calibrating the stylus.The programmer was returned for service.There was no patient involvement.
 
Manufacturer Narrative
Product analysis :analysis was able to confirm the customer comment that the stylus of the programmer was unable to be used, and that the programmer was unable to be calibrated.It was noted that the keyboard and the keyboard rails were broken.It was further noted that the lower case was worn out and the upper and lower display bullets were missing.Reconfigured hard drive and reloaded software.All found defective parts were replaced and all other identified issues were resolved.The programmer then passed final functional and system tests.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK PROGRAMMER
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
MDR Report Key12105892
MDR Text Key259587618
Report Number2182208-2021-02677
Device Sequence Number1
Product Code KRG
UDI-Device Identifier00613994127631
UDI-Public00613994127631
Combination Product (y/n)N
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 07/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090W
Device Catalogue Number2090W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2021
Initial Date Manufacturer Received 06/04/2021
Initial Date FDA Received07/02/2021
Supplement Dates Manufacturer Received07/01/2021
Supplement Dates FDA Received07/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-