Model Number 0998-00-3023-53
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/10/2021
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device: (10/213) a getinge field service engineer (fse) evaluated the iabp, for the reported issue "maintenance code 7" powered the iabp on and observed the unit complete and pass all power on self tests.He connected known system trainer and known balloon and initiated autofill, the device completed autofill and continued pumping at 80bpm for approximately 60 minutes without any alarms or abnormal function occurring.He continued troubleshooting and noticed dust on power supply as well as fault #64 in the logs, cleaned power supply per service manual and completed full pm, safety, calibration, and functionality checks.All checks passed factory specifications.Will allow device to run for 24hrs and follow up with customer to ensure unit has not powered off.On 6-17 followed up with biomed and confirmed device was still running after 24hrs without shutting down.Device is functioning in accordance with factory specifications at this time and is cleared for clinical use.A supplemental report will be submitted upon completion of our investigation.The full name of the initial reporter that is abbreviated is (b)(6).
 
Event Description
It was reported by the customer that prior to in house training the cs300 intra-aortic balloon pump (iabp) had a maintenance error #7 and powered off.There was no patient involvement, and no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12107604
MDR Text Key265584277
Report Number2249723-2021-01427
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 07/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/02/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/11/2004
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
N/A.
-
-