|
As reported by the customer for this event, there was shipment damage to outside of the packaging and also damage to the boxes containing the devices.Total 38 pieces were in the shipment package and all were returned for inspection.Upon visual inspection of the returned devices, five out of the 38 were torn or damaged with holes, thus breaching sterility.This medwatch is being submitted for the fourth device damaged with holes.
|
|
There are five associated medwatch (mdr) reports filed for this event.In the visual inspection of the returned devices damaged in shipment, five out of 38 devices were found to have damage with holes, breaching sterility, for which five mdrs are being submitted.The patient identifiers of the five devices mdrs are: (b)(6) ¿ first device (b)(6)¿ second device (b)(6)¿ third device (b)(6)¿ fourth device (b)(6) ¿ fifth device this mdr is for the fourth device.The customer returned a total of 38 fb-225u devices lot# 05v unused biopsy forceps for evaluation due to shipping damage.The 38 biopsy forceps were received inside the original packages.A visual inspection on the packages was performed, out of the 38 original packages, five packages were torn and damage with holes and scrapes.Out of the five devices, two were severely scraped at the handle area.The 38 devices did not come with the original shipping box.Evaluation is ongoing.Supplemental report(s) will be submitted when any relevant new information is available.
|
|
There is more information on the device evaluation.This supplemental report is being submitted to provide this information.No abnormalities were detected in the device history record (dhr) review, and no similar events have not been reported from other facilities.It can be inferred that a force might have been applied to the sterilization package during the transportation after shipment was made from the factory, or device handling might have contributed to the reported event.The instructions for use includes the following statements: do not store the sterile package containing the instrument in places where it will become damaged, wet, or improperly sealed.Otherwise, the sterility of the instrument may be compromised, which could pose an infection-control risk or cause tissue irritation.Before use, prepare and inspect the instrument as instructed below.Should any irregularity be observed, do not use the instrument; use a spare instead.Damage or irregularity may compromise patient or user safety, for example: posing an infection-control risk; causing tissue irritation, perforation, bleeding, or mucous membrane damage.It may also result in more severe equipment damage.Inspect the sterile package for tears, inadequate sealing, or water damage.If the sterile package shows any irregularities, the sterile condition of the instrument may have been compromised.Do not use the instrument.
|