• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. VISUALASE; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. VISUALASE; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number 9735560
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No parts have been received by the manufacturer for evaluation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding a thermal therapy system being used for a neuro soft tissue ablation procedure.It was reported that intra-operatively, there was a pin hole leak in the saline tubing.The tubing was replaced causing a five minute delay.The procedure was completed and there was no reported impact to patient outcome.
 
Additional Manufacturer Narrative
H3: the catheter was returned to the manufacturer for analysis.Analysis found that the reported issue could not be confirmed; no leak was found during visual inspection.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISUALASE
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key12108782
Report Number1045254-2021-00339
Device Sequence Number19846693
Product Code GEX
UDI-Device Identifier00763000218171
UDI-Public00763000218171
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K053087
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/08/2023
Device Model Number9735560
Device Catalogue Number9735560
Device Lot Number0222215295
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/16/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/01/2021
Supplement Date Received by Manufacturer07/16/2021
Initial Report FDA Received Date07/02/2021
Supplement Report FDA Received Date07/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age39 MO
Patient Weight18
-
-