Model Number TV-AB2980-N
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 06/15/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device involved in this event will not be returned for evaluation as it remains implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.
 
Event or Problem Description
The patient was implanted with the alto stent graft system to treat an abdominal aortic aneurysm (aaa).It was reported that when the polymer was delivered, the contralateral limb did not deploy.All troubleshooting steps were performed; however, the contralateral limb did not deploy so the polymer was not able to fill the limb.Deployment of the ipsilateral limb was completed and the aneurysm was sealed without an endoleak.The physician then performed a femoral-femoral bypass to provide blood flow to both lower extremities.The patient was reported as doing well and both lower extremities have palpable pulses.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.The delivery system was not returned for evaluation.A clinical evaluation of the adverse event/incident could not be completed.No medical records nor medical imaging relevant to the reported adverse event/incident was received by endologix.The user facility was unable to provide this information.Due to the absence of medical records and medical imaging; device, use, procedure, and/or anatomy relatedness to this adverse event/incident could not be evaluated.The final patient status was reported as doing well.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Corrections: g4: date received by manufacturer has been updated.H6: result code: remove code 3233.H6: conclusion code: remove code 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTO
Common Device Name
MAIN BODY
Manufacturer (Section D)
ENDOLOGIX SANTA ROSA
3910 brickway blvd
santa rosa CA 95403
MDR Report Key12110984
Report Number3008011247-2021-00068
Device Sequence Number14602735
Product Code MIH
UDI-Device IdentifierM701TVAB2980N1
UDI-Public+M701TVAB2980N1/$$3240208FS020321332
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P120006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 06/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/08/2024
Device Model NumberTV-AB2980-N
Device Catalogue NumberTV-AB2980-N
Device Lot NumberFS020321-33
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/15/2021
Supplement Date Received by Manufacturer08/03/2021
Initial Report FDA Received Date07/02/2021
Supplement Report FDA Received Date08/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OVATION IX ILIAC LIMB, LOT # FS021921-17; OVATION IX ILIAC LIMB, LOT # FS040621-23; OVATION PRIME FILL POLYMER, LOT # FF011221-01
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
-
-