SMITH & NEPHEW, INC. R3 CONST 54MM; PROSTHESIS, HIP, CONSTRAINED, CEMENTED OR UNCEMENTED, METAL/POLYMER Back to Search Results
Model Number 71339154
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 02/21/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after a tha had been performed on (b)(6) 2021, the patient experienced a dislocation while standing and pivoted.A revision surgery was conducted on (b)(6) 2021 to exchange the oxinium femoral head 12/14 22 mm +0 (case (b)(4)) and the r3 const 54mm (case (b)(4)).Patient outcome is unknown.
 
Additional Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device does not confirm the stated failure mode.The device has deformation along the base which shows signs of impingement with the neck of the stem.There is also gouging along the locking detail likely from extraction.A dimensional inspection was attempted, but the device was too damaged to obtain accurate readings.The clinical/medical evaluation concluded that after three requests no relevant clinical information has been provided for inclusion in this medical investigation.Therefore, a thorough medical investigation cannot be rendered nor can a root cause of the reported dislocation be determined.Based on the information provided, a revision was performed, and the surgeon exchanged the patient¿s oxinium femoral head 12/14 22 mm +0 (case-2021-00061594-1) and the r3 const 54mm (case-2021-00061594-2).Since it was reported, the patient¿s current health status is healthy, no further clinical/medical assessment is warranted at this time.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use documents revealed this failure mode and potential harm was previously identified.Possible causes could include but not limited to traumatic injury, patient anatomy or abnormal loading of limb.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 CONST 54MM
Common Device Name
PROSTHESIS, HIP, CONSTRAINED, CEMENTED OR UNCEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12112154
Report Number1020279-2021-05650
Device Sequence Number14396643
Product Code KWZ
UDI-Device Identifier00885556020227
UDI-Public00885556020227
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
PMA/510(K) Number
K083566
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71339154
Device Catalogue Number71339154
Device Lot Number20GM13959
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2021
Initial Date Received by Manufacturer 06/21/2021
Supplement Date Received by Manufacturer08/16/2021
Initial Report FDA Received Date07/02/2021
Supplement Report FDA Received Date08/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-