MEX03 MEXICO-JUAREZ PRESOURCE KIT, CRANI (RLCU )470; NEUROLOGICAL TRAY Back to Search Results
Model Number PNVXCRLC5
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/14/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
From the device history record, lot# 20201119-23-sh was finished on 12th dec 2020.No exception was recorded in the device history record that could lead to the reported incident.The average linting data is 0.15g/10pcs.No sample returned for investigation.According to the supplier, the operating room towel is made of cotton, so cotton fiber is born.Supplier is continuously working with cardinal health to better control the linting and have implemented several measures to improve it: suctions machines have been installed in grey cloth rolling process, dyeing process and cutting process.The suction process was added before product's final folding, and workers do it according to standard operation procedure requirement.Linting test method and acceptable criteria was stipulated to see the suction results.(b)(4).In the folding process, supplier used one cloth pad under 100 pieces semi-finished products to avoid linting stuck onto the products during product's transfer.From the investigation, no abnormal situation happened in production or dhr.Therefore, the root cause could not be determined.The complaint information was informed to the relevant sectors for their awareness.There is no action taken at this time, but supplier will continue to monitor the trend of this type of incident.
 
Event or Problem Description
Customer reported blue cotton operating room towels pwtb04-stm from the craniotomy pack pnvxcrlc5 are linting causing concerns for surgical site infections.There was no adverse effect to the patient.The scrub and doctors noticed the lint on the instruments, especially when they were working under the microscope.No other information was provided when requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KIT, CRANI (RLCU )470
Common Device Name
NEUROLOGICAL TRAY
Manufacturer (Section D)
MEX03 MEXICO-JUAREZ PRESOURCE
c. arcadia 1580 , col. terraza
cd.juarez , chihuahua CP 32 599
MX  CP 32599
MDR Report Key12112188
Report Number1423537-2021-00646
Device Sequence Number4616035
Product Code OJG
UDI-Device Identifier10195594063788
UDI-Public10195594063788
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPNVXCRLC5
Device Catalogue NumberPNVXCRLC5
Device Lot Number611490
Was Device Available for Evaluation? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/15/2021
Initial Report FDA Received Date07/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-