Model Number SC-1200
Device Problems Use of Device Problem (1670); Wireless Communication Problem (3283)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/09/2021
Event Type  Injury  
Manufacturer Narrative
Date of event: approximated based on the date the manufacturer became aware of the event.
 
Event Description
It was reported that patients ipg would not communicate with the remote control (rc) and suspected to be burned out after the lead revision therefore the physician decided to replaced the ipg.Electrocautery is a known source of high voltage transient signal and current company labeling warns against the use of electrocautery.(physicians implant manual 90970880-04).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION MONTAGE MRI
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key12112632
MDR Text Key260761925
Report Number3006630150-2021-03472
Device Sequence Number1
Product Code LGW
UDI-Device Identifier08714729905943
UDI-Public08714729905943
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 07/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/02/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/09/2021
Device Model NumberSC-1200
Device Catalogue NumberSC-1200
Device Lot Number364379
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/09/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/15/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age70 YR
-
-