| Model Number |
S208 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); High impedance (1291); Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/29/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has been returned for analysis.This report will be updated upon completion of analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that the pacemaker device and right atrial (ra) lead exhibited noise, which was oversensed resulting in inappropriate mode switch episodes.Also, there were high out of range ra lead pace impedance measurements greater than 2500 ohms.The cause of the noise and high impedance measurements is unknown.The boston scientific representative indicated they are not aware of any related patient symptoms.The pacemaker device was explanted, the ra lead was surgically abandoned and both products were replaced.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been returned for analysis.This report will be updated upon completion of analysis.The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.Pin gauge testing, designed to verify proper port dimensions, was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the pacing, and sensing functions were tested.Impedance testing was completed, and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
|
| |
|
Event or Problem Description
|
|
It was reported that the pacemaker device and right atrial (ra) lead exhibited noise, which was oversensed resulting in inappropriate mode switch episodes.Also, there were high out of range ra lead pace impedance measurements greater than 2500 ohms.The cause of the noise and high impedance measurements is unknown.The boston scientific representative indicated they are not aware of any related patient symptoms.The pacemaker device was explanted, the ra lead was surgically abandoned and both products were replaced.No additional adverse patient effects were reported.
|
| |
|