• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GNS II STEMLESS TIB TR SZ 4 RT; PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GNS II STEMLESS TIB TR SZ 4 RT; PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY Back to Search Results
Model Number 71430183
Medical Device Problem Codes Crack (1135); Fitting Problem (2183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/18/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during a tka surgery, the base plate of one (1) gns ii stemless tib tr sz 4 rt chipped and would not fit.Incident occurred with instruments outside the patient.Surgery was resumed, without any delay, using a back-up device.No health consequences were reported.
 
Additional Manufacturer Narrative
H3, h6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device was chipped along the surface, rendering the device inoperative.The device shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GNS II STEMLESS TIB TR SZ 4 RT
Common Device Name
PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12114802
Report Number1020279-2021-05668
Device Sequence Number3624862
Product Code MBH
UDI-Device Identifier03596010211606
UDI-Public03596010211606
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K121393
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71430183
Device Catalogue Number71430183
Device Lot Number14BSA0006
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/09/2021
Initial Date Received by Manufacturer 06/21/2021
Supplement Date Received by Manufacturer07/26/2021
Initial Report FDA Received Date07/05/2021
Supplement Report FDA Received Date07/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-