This follow-up report is being submitted to relay corrected and additional information.Additional: h2, h6.Correction: b4, g3, g6, h10.Event description: it was reported that the patient underwent a initial total hip arthroplasty on (b)(6) 2015 and subsequently the patient was monitored for cobalt levels.The results of cobalt levels test carried on different dates are high.The patient also had concern towards fluid, culture and cell count to be done.The patient underwent multiple mri that detected the possible occurrence of pseudotumor.No mechanical issues were detected with the hip implant and the implant is working fine but the patient is concerned about the elevated cobalt ion levels.A revision surgery has not taken place yet.Review of received data: no medical data relevant to the case has been received.Due diligence: no further due diligence required as all required information to support the conclusion is available or was already requested.Patient data: cobalt level test results were provided by the patient.Review found the results to be increasing in cobalt levels as time progressed.However, medical records for the testing was not provided and thus, could not be reviewed by a health care professional.Product evaluation: no product was returned; therefore, visual and dimensional evaluation could not be performed.Review of product documentation: device purpose: this device is intended for treatment.Product compatibility: the compatibility check was performed from www.Productcompatibility.Zimmer.Com and showed that the product combination was approved by zimmer biomet.Dhr review: review of the device history records identified no deviations or anomalies during manufacturing.Ncr(s): no ncr with a potential correlation to the reported event was found.Raw material certificate: the raw material review shows the device material is conforming to specifications.Conclusion: it was reported that the patient underwent a initial total hip arthroplasty on (b)(6) 2015 and subsequently the patient was monitored for cobalt levels.The results of cobalt levels test carried on different dates are high.The patient also had concern towards fluid, culture and cell count to be done.The patient underwent multiple mri that detected the possible occurrence of pseudotumor.No mechanical issues were detected with the hip implant and the implant is working fine but the patient is concerned about the elevated cobalt ion levels.A revision surgery has not taken place yet.Neither x-rays, operative notes, office visit notes, nor device or photos of the device were received; therefore the condition of the component is unknown.Patient factors that may have affected the performance of the components such as bone quality, activity level, type of activity (low impact vs.High impact), and relevant medical history are unknown.Adherence to rehabilitation protocol is unknown.Based on the investigation the reported event cannot be confirmed.The quality records show that all specified characteristics have met the specifications valid at the time of production.Therefore, the investigation did not identify a nonconformance or a complaint out of box (coob).Based on the given information and the results of the investigation, we were not able to identify a specific root cause for this issue.It is noted that the device is a titanium alloy device, containing no cobalt material.The device can therefore not be connected to the elevated cobalt test levels.However, the root cause of the fluid requiring aspiration cannot be determined.The need for corrective measures is not indicated and zimmer switzerland manufacturing gmbh considers this case as closed.Zimmer biomet¿s reference number of this file is (b)(4).
|