• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MSD BARTLETT MFG CD HORIZON LEGACY 5.5; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MSD BARTLETT MFG CD HORIZON LEGACY 5.5; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 7480150
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/14/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a healthcare provider (hcp) via manufacturer representative regarding a patient with an unknown indication for posterior fusion and decompression with solera instrumentations.It was reported that when using the counter torque the surgeon impacted with a mallet to seat the counter torque and rod to the solera screw.Upon doing the connection between the counter torque and handle fractured on the instrument.Another sets opened to acquire the same instrument to replacement and carried on with the procedure.There was a minor delay in procedure.Products will be returned.There were no patient symptoms or complications reported as a result of this event.Additional information received from manufacturer representative that the counter torque product came in contact with the solera screws which were implanted and attached to the patient.
 
Additional Manufacturer Narrative
Product analysis product id:7480150 lot# ma08f020 visual and optical examination identified that the handle of the counter torque appears to have been sheared off at the welded area of the shaft and is missing.This is consistent with over torque.Additional information: d9, g1, g3, h3, h6.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON LEGACY 5.5
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
MSD BARTLETT MFG
2975 brother blvd
bartlett TN 38133
Manufacturer (Section G)
MSD BARTLETT MFG
2975 brother blvd
bartlett TN 38133
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key12119536
Report Number1030489-2021-00887
Device Sequence Number3490373
Product Code LXH
UDI-Device Identifier00721902762994
UDI-Public00721902762994
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7480150
Device Catalogue Number7480150
Device Lot NumberMA08F020
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/14/2021
Supplement Date Received by Manufacturer12/08/2021
Initial Report FDA Received Date07/06/2021
Supplement Report FDA Received Date01/03/2022
Date Device Manufactured06/01/2008
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexMale
-
-