• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. CATH PKGD: BERMAN 6 FR 90CM; CATHETER, FLOW DIRECTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. CATH PKGD: BERMAN 6 FR 90CM; CATHETER, FLOW DIRECTED Back to Search Results
Model Number IPN000084
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/04/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Qn# (b)(4).
 
Event or Problem Description
It was reported that the md was performing the procedure according to the ifu.The md found that the balloon would not inflate due to a catheter defect.As a result, the md judged it as a defective product and another catheter was used at the same insertion site.There was no report of delay in therapy.There was no report of patient complications, serious injury or death.
 
Event or Problem Description
It was reported that the md was performing the procedure according to the ifu.The md found that the balloon would not inflate due to a catheter defect.As a result, the md judged it as a defective product and another catheter was used at the same insertion site.There was no report of delay in therapy.There was no report of patient complications, serious injury or death.
 
Additional Manufacturer Narrative
Qn# (b)(4).Teleflex received the device for investigation.The reported complaint of balloon would not inflate in use is not confirmed.No damage or abnormalities were noted to the returned sample upon receipt.The balloon inflated as per specifications during the investigation, and the returned device passed visual and functional test specifications.The root cause of the complaint is undetermined.A device history record (dhr) review was not performed.There was no confirmed product failure with the returned sample.Teleflex assessed the risk for the reported complaint.There are no new or revised risk.This will be monitored for any developing trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CATH PKGD: BERMAN 6 FR 90CM
Common Device Name
CATHETER, FLOW DIRECTED
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key12120588
Report Number3010532612-2021-00184
Device Sequence Number19810845
Product Code DYG
UDI-Device Identifier00801902002280
UDI-Public00801902002280
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K892530
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2021
Device Model NumberIPN000084
Device Catalogue NumberAI-07136
Device Lot Number16F19M0098
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2021
Initial Date Received by Manufacturer 06/07/2021
Supplement Date Received by Manufacturer07/27/2021
Initial Report FDA Received Date07/06/2021
Supplement Report FDA Received Date07/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-