| Model Number |
1570-11-150 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Foreign Body Reaction (1868); Pain (1994); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
01/08/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.
|
| |
|
Event or Problem Description
|
|
Asr medical records received.The patient was implanted with asr xl.It was also reported that the patient had a revision surgery however there were no reason or revision notes provided.After review of medical records patient was revised to addressed failed right total hip arthroplasty secondary to mechanical complications of metal on metal hip replacement.Prior to revision patient alleged pain, elevated cobalt and chromium levels.There was some metallosis along the trunnion.Doi: (b)(6) 2007; dor: (b)(6) 2020; right hip.
|
| |
|