| Model Number |
284176 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/08/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Udi: (b)(4).Investigation summary: the complaint device was received at the service center and evaluated.It was reported that the device did not work.Per service reports, this complaint can be confirmed.During the service evaluation the following defects were identified: pedal mechanisms rusted.Sub'd cover oxidized.The pedal mechanisms and sub'd cover were replaced to resolve the issues.After repair, the device was found to be working according to the specifications.The improper maintenance was identified as the root cause for the device failure during the service evaluation.Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|
| |
|
Event or Problem Description
|
|
It was reported by the customer in (b)(6) that preoperatively to an unknown procedure on (b)(6) 2021, it was observed that the foot ped board 4way device did not work.During in-house engineering evaluation, it was determined that the pedal mechanism was corroded.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
|
| |
|