| Model Number |
9734048 |
| Medical Device Problem Codes |
Display or Visual Feedback Problem (1184); Human-Device Interface Problem (2949)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/15/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No patient information provided as no patient was involved when the reported issue occurred.A medtronic representative went to the site to test the equipment.Testing revealed that the reported issue could be replicated.Concomitant medical products: product id: 9 735585, serial/lot #: 3.1.2, ubd: , udi#:.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation device being used outside of a procedure.It was reported that the screen of the navigation system was blue within orthogonal views.It was noted that, while in the planning portion of the application software, the monitor picture was a blue screen instead of exam information which was expected to display the trajectory views, coronal, axial or sagittal views.It was reported that the application menu was visible.There was no patient present when this issue was identified.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: the software team investigated the reported issue and the logs were reviewed but provided insufficient information to determine root cause of the reported behavior.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|